Hach DR6000 UV VIS Spectrophotometer with RFID Technology
庫存單位: LPV441.99.00012
Product Overview
For comprehensive water testing, the DR6000 is your top-tier choice, boasting high-speed wavelength scanning and over 250 pre-programmed methods.
- High-Speed Scanning: Covers UV and visible spectrum.
- Enhanced Accuracy: Utilizes RFID technology and TNTplus test kits.
- Quality Assurance: Integrated software for scheduling and documenting quality measurements.
Product Advantages
- Your Water Testing Needs, All in One Spectrophotometer
The DR6000 has the most pre-programmed testing methods, including high-speed wavelength scanning across the UV and Visible Spectrum. - Accessories for High Volume and High Accuracy Testing Needs
A carousel sample changer allows up to seven sequential measurements. The Sipper Module, an instrument-controlled sample delivery system, increases precision by constant optical characteristics. - Advanced Quality Assurance at Your Fingertips
The DR6000 comes with integrated QA software for scheduling, documenting and interpreting all of your needed quality measurements. - Guided Procedures and Elimination of False Readings
The DR6000, when used with TNTplus® reagent vials, provides the accurate results you need by guiding you step-by-step through your testing procedures. With TNTplus, the instrument averages 10 readings and eliminates outliers, making scratched, flawed or dirty glassware a non-issue. - Automatically Avoids Errors
RFID* technology automatically updates the program calibration factors when you place a TNTplus reagent box near the DR6000. The instrument identifies chemistry expiration dates via a barcode on the vials, and detects chemistry coefficient factors to avoid errors that can occur in lot-to-lot variations in the chemistry.*RFID technology currently available only in US, Anguilla, American Samoa, Australia, Bolivia, Canada, Cayman Islands, Columbia, Dominican Republic, El Salvador, Federated States of Micronesia, Guam, Guatemala, Marshall Islands, New Zealand, Northern Mariana Islands, Palau, Panama, Puerto Rico, and US Virgin Islands.
Brand: Hach
Technical Attributes
| Data Logger | 5000 data points (result, date, time, sample-ID, user-ID) |
| Display | TFT 7 inch WVGA color touch |
| Enclosure Rating | IP20 with closed lid |
| Interfaces | USB type A (2), USB type B, Ethernet, RFID module |
| Manual Languages | en, es, fr, ja, ko, pt, zh |
| Operating Conditions | 10 - 40 °C (50 - 104 °F), max. 80% relative humidity (non-condensing) |
| Operating Mode | Transmittance (%), absorbance and concentration (wavelength, time) |
| Optical System | Reference beam, spectral |
| Photometric Accuracy | 5 mAbs @ 0.0 - 0.5 Abs <1% @ 0.5 - 2.0 Abs @ 546 nm |
| Photometric Linearity | < 0.5 % - 2 Abs |
| Photometric Linearity 2 | 1 % at > 2 Abs with neutral glass at 546 nm |
| Photometric Measuring Range | ± 3 Abs |
| Power Requirements | 100 - 240 V; 50/60 Hz |
| Power Requirements (Hz) | 50/60 Hz |
| Power Requirements (Voltage) | 110 - 240 VAC |
| Preprogrammed Methods | > 240 |
| Sample Cell Compatibility | Rectangular: 10, 20, 30, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch Optional 100 mm rectangular cell with additional adapter |
| Scanning Speed | 900 nm/min (in 1 nm steps) |
| Source Lamp | Tungsten (visible range), deuterium (UV range) |
| Specific Technology | RFID for easy method update, sample ID and Certificate of Analysis |
| Spectral Bandwidth | 2 nm |
| Storage Conditions | -25 - 60 °C (-13 - 140 °F), max. 80% relative humidity (non-condensing) |
| Stray Light | KI-solution at 220 nm < 3.3 Abs/ < 0.05% |
| User Interface Languages | bg, cn, cz, da, en, es, fr, gr, hr, hu, it, jp, kr, nl, pl, pt, ro, ru, sl, sl, sv, tr |
| User Programs | 200 |
| Warranty | 12 months |
| Wavelength Accuracy | ± 1 nm |
| Wavelength Range | 190 - 1100 nm |
| Wavelength Reproducibility | < 0.1 |
| Wavelength Resolution | 0.1 |
| Wavelength Selection | Automatic, based on method selection |
| Weight | 11 kg (24.3 lbs) |
| What's included? | 1x DR6000 UV VIS Spectrophotometer 1x Power Cord (US, EU) 1x Universal-Adapter 1x Dust Cover Matched pair of 1 inch glass sample cells Printed multilingual basic user manual (en, fr, es, pt, zh, jp, ko) |