Jeio Tech Low Temp. Incubators (2/4 chamber)
庫存單位: IL-11-2C / IL-11-4C
Two or four incubation chambers in a single unit.
Product Advantages
Efficient lab space use with multi-chamber incubator
- Designed to offer an ideal and cost-effective solution for laboratories with limited space and budget.
- Two or four 150L chambers merged into a single unit.
- Optimum model selection between two and four chamber unit according to application demand.
- Suitable for conducting each different experiment at once.
- Independent temperature and timer control of each chamber using microprocessor PID.
Performance
- Temperature range from 0℃to 60℃.
- Silent cooling system reduces water evaporation within the chamber.
- Microprocessor PID control.
- Temperature calibration.
- Automatic tuning.
- Adjustable 3-step fan speed with defrosting time and period control.
Convenience
- Intuitive control panel with bright LED display. (0.1℃resolution)
- Dual wait on/off timer modes. (1 min. to 99 hr. 59 min.)
- Convenient preset function for 3 most commonly used temperature settings.
- Up to 9 temperature steps and maximum 200 cycles of repetition programmable for each profile.
- Convenient sample monitoring without affecting chamber temperature via the inner tempered-glass door.
- Rounded inner chamber corners for easy cleaning and optimal airflow.
- Built-in casters for easy transport and installation.
Safety
- Automatic run after power interruption.
- Keypad lock function prevents accidental changes during operation.
- Over-temperature and over-current protection.
- Audible and visible open door alarms.
Please visit the Jeio Tech website for more information
Brand: Jeio Tech
Model: IL-11-2C / IL-11-4C
| Model | IL-11-2C | IL-11-4C |
| Chamber volume (L / cu ft) | 150 / 5.3 x 2 chambers | 150 / 5.3 x 4 chambers |
| Refrigerator (Hp) | 1/8 HBP x 2EA | 1/8 HBP x 4EA |
| Temperature | ||
| Range (℃ / ℉) | 0 to 60℃ / +32 to 140℉ | 0 to 60℃ / +32 to 140℉ |
| 1)Fluctuation at 25℃ (±℃ / ℉) | 0.1 / 0.18 | 0.1 / 0.18 |
| 1)Variation at 25℃ (±℃ / ℉) | 0.5 / 0.9 | 0.5 / 0.9 |
| Dimension(W×D×H) | ||
| Interior (mm / inch) | 600x500x500 / 23.6x19.7x19.7 x 2 chambers |
600x500x500 / 23.6x19.7x19.7 x 4 chambers |
| Exterior (mm / inch) | 760x760x1690/ 29.9x29.9x66.1 | 1500x760x1690/ 59.1x29.9x66.1 |
| No. of wire shelf (standard / max.) | 2 / 11 per chamber | 2 / 11 per chamber |
| Net weight (Kg / lbs) | 180 / 397 | 350 / 772 |
| Electrical requirements (230V, 50Hz) | 9A | 9A x 2EA |
| Cat. No. | AAH28112K | AAH28012K |
1) Technical data according to DIN 12880.
※ Temperature change may occur during automatic defrost.
※ A defrost cycle is applied when operating with a temperature setpoint of 40°C or below.
Temperature and defrost settings may be modified depending on the operating environment. (Request service)
.png)